Nasal sound pressure as objective verification of implant in active transcutaneous bone conduction devices

Journal article
Authors Sabine Reinfeldt
Cristina Rigato
Bo Håkansson
Karl Johan Fredén Jansson
Måns Eeg-Olofsson
Published in Medical Devices: Evidence and Research
Volume 12
Pages 193-202
Publication year 2019
Published at Institute of Clinical Sciences, Department of Otorhinolaryngology
Pages 193-202
Language en
Links doi.org/10.2147/MDER.S197919
Keywords Bone conduction, Bone conduction implant, Ear-canal sound pressure, Nasal sound pressure, Objective intraoperative verification
Subject categories Otorhinolaryngology

Abstract

Objective: Active transcutaneous bone conduction devices consist of an external audio processor and an internal implant under intact skin. During the surgical procedure, it is important to verify the functionality of the implant before the surgical wound is closed. In a clinical study with the new bone conduction implant (BCI), the functionality of the implant was tested with an electric transmission test, where the output was the nasal sound pressure (NSP) recorded in the ipsilateral nostril. The same measurement was performed in all followup visits to monitor the implant’s functionality and transmission to bone over time. The objective of this study was to investigate the validity of the NSP method as a tool to objectively verify the implant’s performance intraoperatively, as well as to follow-up the implant’s performance over time. Design: Thirteen patients with the BCI were included, and the NSP measurement was part of the clinical study protocol. The implant was electrically stimulated with an amplitudemodulated signal generator using a swept sine 0.1-10 kHz. The NSP was measured with a probe tube microphone in the ipsilateral nostril. Results: The NSP during surgery was above the noise floor for most patients within the frequency interval 0.4-5 kHz, showing NSP values for expected normal transmission of a functioning implant. Inter-subject comparison showed large variability, but follow-up results showed only minor variability within each subject. Further investigation showed that the NSP was stable over time. Conclusion: The NSP method is considered applicable to verify the implant’s functionality during and after surgery. Such a method is important for implantable devices, but should be simplified and clinically adapted. Large variations between subjects were found, as well as smaller variability in intra-subject comparisons. As the NSP was found to not change significantly over time, stable transmission to bone, and implant functionality, were indicated.